Pro Lite Technology

Optronik FMS 10 Illuminance Colorimeter

Optronik FMS 10 Illuminance Colorimeter

A laboratory tristmulus colorimeter with research-grade performance

Further Information
» FMS 10 Datasheet » Optronik Photometer Brochure

The Optronik FMS 10 is our research-grade, laboratory tristimulus colorimeter which is available with a choice of sensors which are precision-matched to the CIE X1, X2, Y & Z observer functions. The FMS 10 is a bench-top photometer for measuring both the colour and the illuminance of light sources in units of lux (lx).

Research-Grade Performance & Features

The FMS 10 has a dynamic range of 6 decades, providing measurements up to 600,000 lux with a resolution of 0.0001 lux. An RS-232C (serial) interface is fitted as standard, while optional LightCon FMS software provides PC control, test report generation, data storage and processing, graphical representation, isocolor diagrams and CIE tri-chromatic colour space diagrams. 

Four Filtered Detectors for the Highest Colorimetric Accuracy

The FMS-10 employs four thermally stabilised silicon detectors in its sensor head, each of which is filtered to precisely match a CIE tristimulus XYZ observer function. The use of two detectors for the X function (X1& X2) significantly improves measurement accuracy compared with colorimeters that use only three detectors in total. Optronik developed one of the first digital lux meters in 1968 and is today acknowledged as a leader in the design and manufacture of precision photometric equipment. The proprietary tristimulus response filters consist of several elements designed to match the standard observer functions. The f1' figure-of-merit (for the photopic, Y channel) is better than 1.5% at all wavelengths (for class L sensors per DIN 5032) or better than 2.0% (for the class A sensor). The out-of-band sensitivity in the IR and UV range is also reduced to below 0.1%. The careful design of the detectors ensures best-in-class accuracy and repeatability, even with monochromatic or narrow band radiation sources such as LEDs. The FMS 10 is available with a choice of 3 standard sensors with active areas of 14 and 60mm diameter. Depending upon the sensor, you can choose either a DIN class A (f1' < 2.0%) or class L (f1' < 1.5%) spectral response. The 14mm sensor was developed for improved accuracy with highly directional light sources with uneven light distributions. The smaller sensor area helps to ensure that the detector is fully illuminated even where the device under test has a narrow beam angle.

Specifications: FMS 10 Meter

Model Selection

Sensor Type

FMS 10-14L

CE 10-14L sensor (14mm, class L)

FMS 10-60A

CE 10-60A sensor (60mm, class A)

FMS 10-60L

CE 10-60L sensor (60mm, class L)

Specifications

For sensor specifications, see below

Measurement Units

lux (lx)

Option: candelas (cd) or lumens (lm)

 

X Y Z (CIE-LAB), x y Y (CIE-LAB),
u v Y (CIE-LAB), u' v' Y (CIE-UCS 1976)

 

Tcp (Kelvin)

Dynamic Range

6 decades

Automatic range selection

Measurement Range

High: 600,000 lx

Low: 2 lx (with 0.0001 lx resolution)

Display

Graphical 4-digit LCD with backlight

 

Representation of CIE colour space diagram

Digital Interface

RS-232C (serial)

Compliance 

DIN 5032 (class A or L); CIE No. 69

Dimensions (W x H x D)

450 x 325 x 135mm 

Housing

19-inch rack mountable (3U)

Option: benchtop housing with handle

Weight

8 kg

Power

230 ± 10% Volts, 50Hz

 

< 50 VA consumption

Options
Display of alternative units (e.g. lumens)
PTB calibration certificate
Stray light tube & tridpod (for specific working distance/s)
Extended cable length between meter and sensor (5, 10, 15, 20 or 25m)
LightCon control software
Benchtop type housing with carrying handle

Specifications: CE-10 Series Sensors

CE 10-

14L

60A

60L

Type

Four silicon photodiodes

Aperture (mm)

14

60

60

V(l) f1' (%)

< 1.5

< 2.0

< 1.5

UV & IR sensitivity (%)

< 0.1

Cosine Correction

No

Thermal Stabilisation

Yes

Thermal Drift a0 (%/K)

<0.01

DIN 5032 Class (Y channel)

L

A

L

Ev (min) (mlx)

1

Ev (max) (klx)

600

Calibration

Against luminous intensity standard lamp with
calibration certificate from PTB (Physikalisch-
Technische Bundesanstalt, Germany)

Uncertainty (%)

± 0.6 traceable to NIST

Dimensions (mm)

ø98 x 165

Weight (kg)

0.9

Cable Length (m)

3 (cable extensions optional - see above)

Data Supplied 
Individual test report for spectral sensitivity
Specified errors per to DIN 5032 § 6 & 7 and CIE Publ. No. 69 (1987)
Optronik calibration certificate

Related Products
Datasheets
Brochures
Case Studies
Knowledge Base
Have a Question?

Why not use our convenient enquiry form?

> Click Here

Buy Online

This product can be ordered via the Pro-Lite on-line shop.

> Click Here

Need a measurement?

Contract measurement services are available from Pro-Lite Photometrics using this product

> Click Here