USB2000+IRRAD Irradiance Spectroradiometer
The USB2000+IRRAD is Pro-Lite's modular spectroradiometer system based upon the Ocean Optics USB2000+ spectrometer. This system has the flexibility to measure spectral radiant flux, intensity & irradiance as well as derived photometric (illuminance, luminous intensity) & colorimetric (chromaticity, colour temperature & colour rendering) properties. Using interchangeable collection optics & integrating spheres, you can configure - and reconfigure - your light measurement system for measuring solar irradiance as well as the output of lamps, LEDs, optical fibres, lasers - in fact almost any type of light source. Configure & Reconfigure Your System for Irradiance, Intensity & FluxThe USB2000+IRRAD spectroradiometer can be user configured and reconfigured for measuring spectral irradiance (W/m2.nm), spectral radiance intensity (W/sr.nm) and spectral radiant flux (W/nm) in the 350-1000nm spectral region. To convert from one measurement geometry to another, just select the appropriate collection optics. For irradiance and intensity, you attach a cosine receiver (often called a diffuser) to the input of the optical fibre, while for total flux you simply unscrew the cosine receiver and attach the fibre to one of our integrating spheres. The diffuser option provides a cosine corrected collection of light over a complete hemisphere for accurate irradiance measurements, while an integrating sphere captures the total light emitted from a lamp or LED.
|
| Model |
USB2000+IRRAD Spectroradiometer |
| Product Code | PLT/USB2000+IRRAD |
| Spectral Range | 350-1000nm |
| Spectral Resolution | 1.5nm |
| System Configuration | USB2000+ spectrometer w/grating, slit, collection lens & order sorting filter |
| Optical fibre, 2m long |
|
| Collection Optics | CC-3-UV-S cosine collector |
| FOIS-1 integrating sphere | |
| LMS-100 integrating sphere | |
| Calibration Sources | LS-1-CAL (for CC-3-UV-S) |
| LS-1-CAL-INT (for FOIS-1) | |
| SCL-600 (for LMS-100) | |
| System Software |
SpectraSuite |
| Measurement Capabilities: |
|
| - with CC3-UV-S cosine collector | Spectral irradiance, 350-1000nm (W/m2.nm) |
| Spectral radiant intensity, 350-1000nm (W/sr.nm) | |
| Illuminance (lux) | |
| Luminous intensity (candelas) | |
| - with FOIS-1 integrating sphere |
Spectral radiant flux (forward), 350-1000nm (W/nm) |
| Forward luminous flux (lumens) | |
| - with LMS-100 integrating sphere | Spectral radiant flux (forward & total), 350-1000nm (W/nm) |
| Forward & total luminous flux (lumens) | |
| Colorimetric Capabilities: |
CIE tristimulus coefficients (XYZ) |
| CIE 1931 chromaticity coordinates (xy) | |
| CIE 1976 chromaticity coordinates (u'v') | |
| Correlated colour temperature, CCT (Kelvin) | |
| Colour rendering index, Ra, R1-14 | |
| Dominant wavelength & purity | |
| Optional Accessories | OmniDriver developers software |
| Auxiliary correction facility on integrating sphere |
Why not use our convenient enquiry form?
This product can be ordered via the Pro-Lite on-line shop.
Measurement services are available from Pro-Lite using this product